-40%
742 DIRECT-READING DOSIMETER, ARROW-TECH - 742, 0-200R
$ 42.21
- Description
- Size Guide
Description
New old stock military surplus.Direct-Reading Dosimeter - 742, 0-200R
-742 dosimeter
-Cert of calibration 06/13
-Fluke Biomedical Cd: Nuclear Associates Direct Reading Dosimeters: Manual/Warranty/Data Sheet
Direct-Reading Dosimeter with Sapphire Window
Model 742, 0-200R
The Direct-Reading Dosimeter is pocket-size, carbon-fiber electroscope with an ion chamber for detecting and indication of integrated exposure to Gamma and X-Ray radiation. It has a thin wall which permits the penetration and detection of radiation.
Dosimeters are extremely sensitive instruments. Although they are constructed for rugged use, they should receive the same care as a wristwatch. Since dosimeters are hermetically sealed at the factory, they cannot be repaired or adjusted in the field. Therefore, if instrument malfunctioning is experienced, the instrument should be returned to the factory or your dealer. Dosimeters may be maintained in operating condition simply by cleaning the eyepiece lens and the charging switch insulator with clean water and a cloth that is free of lint and grit. Make sure the charging switch insulator is absolutely free of lint and moisture at all times. Caution: Do not insert any sharp objects into, or tamper with parts in the charging switch recess. Irreparable damage may be done.
Arrow-Tech, Inc. recommends a yearly calibration of the Direct-Reading Dosimeter which is generally consistent with good health physics practices. More frequent calibration may be necessary should the user's license require a shorter calibration interval.
Licensed from the U.S. Government under Patent No. 5,426,305
Specifications:
Radiation Detected:
Gamma and X-Ray from 16 keV to 6 MeV
Range:
0-200R
Detector:
Fiber-Electrometer mounted in an Electrically Conductive Plastic Ion Chamber
Detector Housing:
V
ery low permeability plastics - Hermetically-Sealed
Accuracy:
+/– 10% of true exposure
Rate Response:
Dose rate independent for Gamma and X-Ray
Electrical Leakage:
Less than 1.0% of full scale for 24 hours at 50 C
Temperature Range:
-20 degrees C to +50 degrees C
Relative Humidity:
Up to 90%
Dimensions:
4.5” (12.4 cm) Length X .6“ (1.5 cm) Diameter
Weight:
1.0 oz. (25 grams)
Finish:
Natural Matte Gray with a Metal Clip
Warranty:
Two Year Limited Warranty
Shelf Life:
20 Years
Specifications:
ANSI N13.5 and ANSI N322-1997